Phase Contrast

When performing in-line (propagation-based) phase-contrast X-ray tomographic measurements, images are obtained with both absorption and refraction contribution. This allows to obtain an enhanced contrast (Fresnel diffraction) at the edges of sample features producing local phase variations. By this technique, 3D variations of the electron density in the sample can be resolved and the experiments aim at a quantitative 3D reconstruction of the complex refraction index. The recorded sample radiographs are determined by the projection of the real and imaginary part of the refractive index within the sample volume. The Fresnel fringes visible at the edges of the phase objects can be considered as an artefact in the reconstructed slices and should be reduced or removed before any quantitative post-processing and analysis of the images.

Radiographs can be recorded at a single or multiple distances. If a single distance is employed, a phase retrieval algorithm is applied most often based on the homogeneity assumption and the transport of intensity equation proposed by [B35]. In the case of multiple distances [B8], the problem is afforded by using contrast transfer functions. Because the phase-retrieval application usually produces a blurring of the reconstructed slices, it is crucial to optimize the algorithm parameters in order to obtain a good compromise between spatial resolution and discrimination of the phases of interest in the final images.

The common reconstruction approach foresees a two-steps procedure: 1) a phase-retrieval algorithm is individually applied to the acquired projections; 2) a full tomographic reconstruction is then performed using as input data the phase-retrieved projections.

Multi-distance

Discreteness in propagation based (multi-distance) phase contrast tomography [B18]

Intrinsic limitations in recovering low spatial frequencies in the phase retrieval step usually lead to fluctuating background in the tomographic reconstructions (Figure 1, right panel). In particular in the case of weak contrast between the constituent phases of the material, the 3D segmentation of these different regions may no longer be possible. In the case of objects with discrete values of the refractive index, combined phase retrieval & reconstruction schemes like the one pioneered in [B23] may allow solving this problem. Further research and developments in this direction are required.

The challange of this dataset includes incorporate / impose / use knowledge about discreteness of the refractive index inside an object in order to improve the quality of tomographic reconstructions of the refractive index from (multi-distance) propagation based phase contrast tomography.

These multi-distance phase contrastat datasets contains 4 measurements of the same sample a Duplex (austenitic, ferritic) steel at 4 sample-to-detector distances (8, 58, 158 and 308 mm). The two sample constituents phases austenite (fcc) and ferrite (bcc) only have a weak difference in electron density. The projection images also contains dark diffraction contrast spot (some grains fulfill the Bragg condition and give rise to diffracted beams). These contrast appear and disappear from projection to projection and act as an additional perturbation/noise. The experimental conditions reported in the table below:

Instrument

ESRF

Energy

55 keV

Scan Range

360 degree

Number of Projections

450

White Fields

21

Dark Fields

21

Ref ON

150

Rotation Speed

0.75 deg/s

Sample Detector Distance

8, 58, 158 and 308 mm

Exposure Time

2 s

Pixel Size

1.4 µm

Detector Dimension x

350

Detector Dimension y

650

The sphere data sets includes 4 tomographic data sets collected at different sample to detector distance and different concentrations as reported in the table below:

To load the data sets and perform a basic reconstruction using tomopy

tomopy recon --file-name tomo_00064.h5 --rotation-axis 175
tomopy recon --file-name tomo_00065.h5 --rotation-axis 175
tomopy recon --file-name tomo_00066.h5 --rotation-axis 175
tomopy recon --file-name tomo_00067.h5 --rotation-axis 175

To enable phase retrieval un-comment the appropriate setting in tomopy_rec.py

Tomo ID

Distance (mm)

Sample Name

Image

Axis

tomo_00064

8

p2_pct_1

00064

175

tomo_00065

58

p2_pct_2

00065

175

tomo_00066

158

p2_pct_3

00066

175

tomo_00067

308

p2_pct_4

00067

175

Wet Sample

The evaluation of wettability and swelling properties of biomaterials is fundamental in several applications in order to assess the structural stability as well as the ability of water uptake of the synthesized 3D structures [B2]. However, experiments carried out in wet conditions might be affected by motion artfacts. These artifacts negatively affect image interpretation.

Here we present a dataset in which the challange is to compensate a partial motion artifact occurring in air bubbles within a liquid-embedded biomaterial structure causing the edges of the air bubbles to contain streaks due to motio of the air bubbles instead of forming a complete contour. The experimental conditions reported in the table below:

To load the data sets and perform a basic reconstruction using tomopy

tomopy recon --file-name tomo_00068.h5 --rotation-axis 1463.5

tomo_ID

00068

Image preview

00068

Downloads

tomo_00068

Instrument

Elettra Syrmep

Sample name

wet sample

X-ray energy

14 keV

Sample-to-detector distance

150 mm

Pixel size

4.1 µm

Scan Range

180 degree

Number of Projections

1800

Rotation axis location

1463.5