vmmosscience - Extraction of scientific spectra
esorex vmmosscience [OPTIONS] FILE.sof
This recipe is used to reduce scientific spectra using the
extraction mask and the products created by the recipe vimos_calib. The
spectra are bias subtracted, flat fielded (if a normalised flat field is
specified) and remapped eliminating the optical distortions. The wavelength
calibration can be optionally upgraded using a number of sky lines: if no
sky lines catalog of wavelengths is specified, an internal one is used
instead.
If the alignment to the sky lines is performed, the input
dispersion coefficients table is upgraded and saved to disk, and a new CCD
wavelengths map is created. A configuration table (typically depending on
the grism in use) may also be specified: this table contains a default
recipe parameter setting to control the way spectra are extracted for a
specific instrument mode, as it is used for automatic run of the pipeline on
Paranal and in Garching. If this table is specified, it will modify the
default recipe parameter setting, with the exception of those parameters
which have been explicitly modifyed on the command line. If a configuration
table is not specified, the input recipe parameters values will always be
read from the command line, or from an esorex configuration file if present,
or from their generic default values (which are rarely meaningful).
Either a scientific or a standard star exposure can be specified
in input.
Only in case of a standard star exposure input, the atmospheric
extinction table and a table with the physical fluxes of the observed
standard star must be specified in input, and a spectro-photometric table is
created in output. This table can then be input again to this recipe, always
with an atmospheric extinction table, and if a photometric calibration is
requested then flux calibrated spectra (in units of erg/cm/cm/s/Angstrom)
are also written in output.
Input files:
DO category: Type: Explanation: Required:
MOS_SCIENCE Raw Scientific exposure Y
or MOS_STANDARD Raw Standard star exposure Y
MASTER_BIAS Calib Master bias Y
SKY_LINE_CATALOG Calib Sky lines catalog .
MOS_MASTER_SCREEN_FLAT Calib Normalised flat field .
MOS_DISP_COEFF Calib Inverse dispersion Y
MOS_CURV_COEFF Calib Spectral curvature Y
MOS_SLIT_LOCATION Calib Slits positions table Y
CONFIG_TABLE Calib Configuration table .
In case MOS_STANDARD is specified in input,
EXTINCT_TABLE Calib Atmospheric extinction Y
STD_FLUX_TABLE Calib Standard star flux Y
TELLURIC_CONTAMINATION Calib Telluric regions list .
In case a photometric calibration is requested for scientific
data, the following inputs are mandatory:
EXTINCT_TABLE Calib Atmospheric extinction Y
MOS_SPECPHOT_TABLE Calib Response curves Y
If requested for standard star data, the SPECPHOT_TABLE can be dropped:
in this case the correction is applied using the SPECPHOT_TABLE produced
in the same run.
Output files (acronym _STD_ replaces _SCI_ and _STANDARD_ replaces
_SCIENCE_ in case input is MOS_STANDARD rather than MOS_SCIENCE):
DO category: Data type: Explanation:
MOS_SCIENCE_REDUCED Image Extracted scientific spectra
MOS_SCI_SKY_REDUCED Image Extracted sky spectra
MOS_SCI_ERROR_REDUCED Image Errors on extracted spectra
MOS_UNMAPPED_SCIENCE Image Sky subtracted scientific spectra
MOS_SCIENCE_EXTRACTED Image Rectified scientific spectra
MOS_SCIENCE_SKY_EXTRACTED Image Rectified science spectra with sky
MOS_SCIENCE_SKY Image Rectified sky spectra
MOS_SCI_UNMAPPED_SKY Image Sky on CCD
MOS_SCI_GLOBAL_SKY_SPECTRUM Table Global sky spectrum
OBJECT_SCI_TABLE Table Positions of detected objects
Only if fringing correction is requested (dithered exposures):
MOS_SCI_FRINGES Image Fringe map
Only if the sky-alignment of the wavelength solution is requested:
MOS_SCI_SKYLINES_OFFSETS_SLIT Table Sky lines offsets
MOS_SCI_DISP_COEFF_SKY Table Upgraded dispersion coefficients
MOS_SCI_WAVELENGTH_MAP_SKY Image Upgraded wavelength map
Only if a MOS_STANDARD is specified in input:
MOS_SPECPHOT_TABLE Table Efficiency and response curves
Only if MOS_SPECPHOT_TABLE or MOS_MASTER_RESPONSE are specified in input:
MOS_SCIENCE_FLUX_REDUCED Image Flux calibrated scientific spectra
MOS_SCI_ERROR_FLUX_REDUCED Image Errors on flux calibrated spectra
MOS_SCIENCE_FLUX_EXTRACTED Image Flux calibrated slit spectra
- --skyalign
<int>
- Polynomial order for sky lines alignment, or -1 to avoid alignment (int;
default: -1). The full name of this option for the EsoRex configuration
file is vimos.vmmosscience.skyalign [default = -1].
- --flatfield
<bool>
- Apply flat field (bool; default: True). The full name of this option for
the EsoRex configuration file is vimos.vmmosscience.flatfield
[default = True].
- --skyglobal
<bool>
- Subtract global sky spectrum from CCD (bool; default: False). The full
name of this option for the EsoRex configuration file is
vimos.vmmosscience.skyglobal [default = False].
- --skymedian
<bool>
- Sky subtraction from extracted slit spectra (bool; default: False). The
full name of this option for the EsoRex configuration file is
vimos.vmmosscience.skymedian [default = False].
- --skylocal
<bool>
- Sky subtraction from CCD slit spectra (bool; default: True). The full name
of this option for the EsoRex configuration file is
vimos.vmmosscience.skylocal [default = True].
- --cosmics
<bool>
- Eliminate cosmic rays hits (only if global or local sky subtraction is
also requested) (bool; default: True). The full name of this option for
the EsoRex configuration file is vimos.vmmosscience.cosmics
[default = True].
- --slit_margin
<int>
- Number of pixels to exclude at each slit in object detection and
extraction (int; default: 3). The full name of this option for the EsoRex
configuration file is vimos.vmmosscience.slit_margin [default =
3].
- --ext_radius
<int>
- Maximum extraction radius for detected objects (pixel) (int; default: 6).
The full name of this option for the EsoRex configuration file is
vimos.vmmosscience.ext_radius [default = 6].
- --cont_radius
<int>
- Minimum distance at which two objects of equal luminosity do not
contaminate each other (pixel) (int; default: 0). The full name of this
option for the EsoRex configuration file is
vimos.vmmosscience.cont_radius [default = 0].
- --ext_mode
<int>
- Object extraction method: 0 = aperture, 1 = Horne optimal extraction (int;
default: 1). The full name of this option for the EsoRex configuration
file is vimos.vmmosscience.ext_mode [default = 1].
- --detection
<float>
- Object detection threshold (ADU) (float; default: 2.0). The full name of
this option for the EsoRex configuration file is
vimos.vmmosscience.detection [default = 2.0].
- --resp_fit_nknots
<int>
- Number of knots in spline fitting of the instrument response. (-1: No
fitting. -2: Read from grism table) (int; default: -2). The full name of
this option for the EsoRex configuration file is
vimos.vmmosscience.resp_fit_nknots [default = -2].
- --resp_fit_degree
<int>
- Degree of polynomial in fitting of the instrument response. (-1: No
fitting. -2: Read from grism table) (int; default: -2). The full name of
this option for the EsoRex configuration file is
vimos.vmmosscience.resp_fit_degree [default = -2].
- --resp_ignore_mode
<str>
- Types of lines/regions to ignore in response. Valid ones are
´stellar_absorption´, ´telluric´ and
´command_line´ (from parameter resp_ignore_lines) (str;
default: ´stellar_absorption,telluric,command_line´). The
full name of this option for the EsoRex configuration file is
vimos.vmmosscience.resp_ignore_mode [default =
stellar_absorption,telluric,command_line].
- --resp_ignore_points
<str>
- Extra lines/regions to ignore in response. Use a comma separated list of
values. A range can be specified like 4500.0-4600.0 (str; default:
´´). The full name of this option for the EsoRex
configuration file is vimos.vmmosscience.resp_ignore_points
[default = ].
- --resp_use_flat_sed
<str>
- Use the flat SED to normalise the observed spectra. Value are true, false,
grism_table. (str; default: ´grism_table´). The full name of
this option for the EsoRex configuration file is
vimos.vmmosscience.resp_use_flat_sed [default =
grism_table].
- --resp_shift
<float>
- The extracted standard star will be shifted these many angstroms before
using it to compute the response. This is useful for observed std stars
not centered in the slits. Positive values will shift the spectrum to the
red. Shift is given in Angstroms but no fraction of pixels will be
shifted. (float; default: 0.0). The full name of this option for the
EsoRex configuration file is vimos.vmmosscience.resp_shift [default
= 0.0].
- --time_normalise
<bool>
- Normalise output spectra by the exposure time (bool; default: True). The
full name of this option for the EsoRex configuration file is
vimos.vmmosscience.time_normalise [default = True].
- --anyframe
<bool>
- Look for a standard star in any frame classified as MOS_STANDARD (bool;
default: False). The full name of this option for the EsoRex configuration
file is vimos.vmmosscience.anyframe [default = False].
- --alignment
<str>
- Type of alignment of dithered frames, either to the nearest neighbour
pixel or to fractions of pixel (str; default: ´integer´).
The full name of this option for the EsoRex configuration file is
vimos.vmmosscience.alignment [default = integer].
- --stack_method
<str>
- Frames combination method (str; default: ´average´). The
full name of this option for the EsoRex configuration file is
vimos.vmmosscience.stack_method [default = average].
- --minrejection
<int>
- Number of lowest values to be rejected (int; default: 1). The full name of
this option for the EsoRex configuration file is
vimos.vmmosscience.minrejection [default = 1].
- --maxrejection
<int>
- Number of highest values to be rejected (int; default: 1). The full name
of this option for the EsoRex configuration file is
vimos.vmmosscience.maxrejection [default = 1].
- --klow
<float>
- Low threshold in ksigma method (float; default: 3.0). The full name of
this option for the EsoRex configuration file is
vimos.vmmosscience.klow [default = 3.0].
- --khigh
<float>
- High threshold in ksigma method (float; default: 3.0). The full name of
this option for the EsoRex configuration file is
vimos.vmmosscience.khigh [default = 3.0].
- --kiter
<int>
- Max number of iterations in ksigma method (int; default: 999). The full
name of this option for the EsoRex configuration file is
vimos.vmmosscience.kiter [default = 999].
- --dither
<bool>
- Align dithered frames before stacking(for multiple input frames) (bool;
default: True). The full name of this option for the EsoRex configuration
file is vimos.vmmosscience.dither [default = True].
- --compute
<bool>
- Compute offsets of dithered images from detected objects (true), or read
offsets from header (false) (bool; default: False). The full name of this
option for the EsoRex configuration file is
vimos.vmmosscience.dither.compute [default = False].
- --fringing
<bool>
- Apply fringing correction (only for dithered observations) (bool; default:
True). The full name of this option for the EsoRex configuration file is
vimos.vmmosscience.fringing [default = True].
- --offset
<float>
- Minimum required offset between exposures for applying the sky fringing
correction. (float; default: 3.0). The full name of this option for the
EsoRex configuration file is vimos.vmmosscience.fringing.offset
[default = 3.0].
- --generate_idp
<bool>
- Generate IDP products (bool; default: False). The full name of this option
for the EsoRex configuration file is
vimos.vmmosscience.generate_idp [default = False].
Note that it is possible to create a configuration file containing
these options, along with suitable default values. Please refer to the
details provided by the 'esorex --help' command.
The full documentation for the vimos pipeline can be downloaded as
a PDF file using the following URL:
- ftp://ftp.eso.org/pub/dfs/pipelines/vimos/vimos-pipeline-manual-6.9.pdf
An overview over the existing ESO pipelines can be found on the
web page https://www.eso.org/sci/software/pipelines/.
Basic documentation about the EsoRex program can be found at the
esorex (1) man page.
It is possible to call the pipelines from python using the
python-cpl package. See
https://packages.python.org/python-cpl/index.html for further
information.
The other recipes of the vimos pipeline are
vimos_ima_bias(7), vimos_ima_dark(7),
vimos_ima_det_noise(7), vimos_ima_fringe(7),
vimos_ima_science(7), vimos_ima_standard(7),
vimos_ima_twilight_flat(7), vmbias(7), vmdark(7),
vmdet(7), vmifucalib(7), vmifucombine(7),
vmifucombinecube(7), vmifuscience(7), vmifustandard(7),
vmmoscalib(7), vmspphot(7)
Carlo Izzo <https://support.eso.org>
Please report any problems to https://support.eso.org.
Alternatively, you may send a report to the ESO User Support Department
<usd-help@eso.org>.
This file is part of the VIMOS Instrument Pipeline Copyright (C)
2019 European Southern Observatory
This program is free software; you can redistribute it and/or
modify it under the terms of the GNU General Public License as published by
the Free Software Foundation; either version 2 of the License, or (at your
option) any later version.
This program is distributed in the hope that it will be useful,
but WITHOUT ANY WARRANTY; without even the implied warranty of
MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General
Public License for more details.
You should have received a copy of the GNU General Public License
along with this program; if not, write to the Free Software Foundation,
Inc., 59 Temple Place, Suite 330, Boston, MA 02111-1307 USA